Specs/Technical
PHYSICAL SPECIFICATIONS
Unit Dimensions: Height 44cm Width 36 cm Length 58cm
Unit Weight: 11 Kg (Excluding Soda lime canister and Vapouriser)
18 Kg (Including Soda Lime canister and Vapouriser)
Materials: High impact plastic polymer body. Anodised Aluminium fittings
ENVIRONMENTAL REQUIREMENTS
OPERATION (Including Differential pressure sensors)
Temperature: 0 - 40 degrees C
Humidity: 15 - 95% relative (Non-Condensing)
Altitude: 500 to 800 mmHg (-440 to 3565 meters).
STORAGE (Including Differential pressure sensors)
Temperature: -25 - 65 degrees C
Humidity: 15 - 95% relative (Non-Condensing)
Altitude: 500 to 800 mmHg (-440 to 3565 meters).
ELECTROMAGNETIC COMPATIBILITY
Immunity: Complies with requirements of EN60601-1-2
Emissions: Complies to CISPR 11
Compliance: CE (pending)
UL60601
IEC 62133
BREATHING SYSTEM SPECIFICATIONS
OPERATIONAL
Breathing system is Circle Mode or attached Ayres –T- piece
CARBON DIOXIDE ABSORBENT CANISTER
Capacity: 500 - 1000 g. Reusable or disposable type
PORTS AND CONNECTIONS
Exhalation: 22mm OD ISO Taper (RED)
Inspiration: 22 OD, 15 ID mm ISO Taper (GREEN)
Self Inflating Bag Inlet: 22mm ID
PRESSURE GAUGE (patient airway pressure)
Scale Range: -20 to 80 cm H2O ( -2 to 7.85 kPa)
ANAESTHETIC GAS SCAVENGING
30mm OD ISO Taper
PNEUMATIC SPECIFICATIONS
GAS SUPPLY
Options: O2 Pipeline, Cylinder or O2 concentrator
Air Pipeline, Cylinder or Air Compressor
Pipeline: 280 to 600 kPa ( 41 - 87psi )
Connections: DISS-male, DISS-female,
ISO 5359, AS4059, FS90-116
Cylinder Input: Pin Index preferred, DIN
PRESSURE GAUGE (gas supply)
Oxygen: 0 - 600 kPa ( 0 - 87 psi )
Air: 0 - 600 kPa ( 0 - 87 psi )
FLOWMETERS
Oxygen: 0 - 8 Litres / Min
Air 0 - 8 Litres / Min
O2 FLUSH 35 to 75 Litres min
VAPOURISER CONNECTION
Type: Selectatec back bar Plenum
Mount: Variable and lock height adjustment
ELECTRICAL SPECIFICATIONS
POWER INPUT
External: Sealed Lead Acid Smart Charger DC
Internal: Sealed Lead Acid Battery 12 volt 2.3 Ah rechargeable
Fuses: Fuse ratings are 2A and 3.15A. Part numbers: S500-2-R and S500-3.15-R
VENTILATOR SPECIFICATION
CONTROL INTERFACE SPECIFICATION
On / Off button: Push On/off toggle type (GREEN: ACTIVE)
Battery Condition: Descending 4 Range LED
I:E ratio: Rotary
Range: OFF - 1:1,1:2,1:3 1:4
Breaths/Min: Rotary
Range: 5 - 30 B/M
Tidal Volume (TV) Rotary
Variable to 1200 ml
ALARM SETTINGS
LOW OXYGEN SUPPLY PRESSURE
Pressure range activation : < 50% Initial Pressure (within preset limit)
(upper limit detect)
PATIENT AIRWAY PRESSURE
High: > 50 cm H2O (alarms and terminates cycle for one breath)
Low: < 5cm H2O above PEEP
ALARM MUTE
Single Press: 30 secs
Press and hold for 2 secs. 90 secs
FOR INFORMATION ON THE HYVAN TRANSPORT BOX, MONITORS AND VAPORIZERS - SEE RESOURCES SECTION
Product introduction:
The HYV 4100 series multigas analyser is the ideal low sampling-rate anaesthesia solution for your patient monitoring system. It can show an accurate gas concentration with all respiration rates up to 150 breaths per minute. Advanced technology enables HYV 4100 to provide accurate measurement of CO2, NO2, anaesthetic agents gas, respiration rate, inspiratory time and expiratory time. With a sampling-rate low to 50 mL/min, HYV 4100 is ideal for use on adult, paediatric and infant patients; patients undergoing procedural sedation; and short- term monitoring of intubated patients.
This device clips onto the side of the Hyvan Monitor and the display can be read on the monitor screen
Product features:
Automatic Agent Identification
Technical Specifications:
Transducer Type |
HYV 4100 series multigas analyser |
||
Sample Rate |
50mL/min,±10mL/min |
||
Principle of Operation |
Non-dispersive infrared(NDIR),no moving parts (Excluding pump) |
||
Initialization Time |
20 sec, full specifications within 60sec |
||
Calibration |
No routine user calibration required |
||
Compensation |
Automatic for atmospheric pressure, temperature |
||
Rise Time |
CO2<200ms, N2O,AA<350ms |
||
Respiratory Rate |
Range:3-150 BPM Accuracy ±1BPM |
||
Breath Detect |
Adaptive threshold, minimum 1% CO2 change |
||
Agent Threshold |
Agent 0.15vol% |
||
Measurement Accuracy |
Gas Type |
Range |
Accuracy |
CO2 |
0-10% |
±(0.2vol%+2% of reading) |
|
|
|
10-15% |
±(0.2vol%+2% of reading) |
N2O |
0-100% |
±(2vol%+2% of reading) |
|
ISO |
0-6% |
±(0.2vol%+2% of reading) |
|
ENF |
0-6% |
±(0.2vol%+2% of reading) |
|
SEV |
0-8% |
±(0.2vol%+2% of reading) |
|
Interface |
RS-232 Operating at 9600bps |
||
Voltage Requirements |
5.0DV±5% |
||
Power Consumption |
Less than 2.0 Watts typical |
||
Temperature and Humidity |
Operating:0-40ºC,10-90% RH, non-condensing Storage: -40-70ºC,<90%RH, non-condensing |
||
Size |
96*57.5*30mm |
||
Weight |
110g |